Sterility Isolators
A quick-acting airlock allows easy loading and unloading of materials, while a fully automated H₂O₂-decontamination process (LOG6 germ reduction) guarantees reliable sterilisation before testing. The system is intuitively controlled via touchscreen and features comprehensive environmental monitoring.




Key Features & Benefits
Designed for maximum sterility assurance, our Sterile Test Isolators from Ortner combine advanced decontamination technology, comprehensive environmental monitoring, and ergonomic operation to ensure safe and reliable sterility testing.
- Integrated H₂O₂ gas generator for stable & reproducible decontamination
- Continuous tracking of airflow velocity, pressure, temperature, and humidity
- Real-time 0.5 µm & 5.0 µm particle monitoring for sterility assurance
- Quick-lock airlocks allow safe inward/outward transfer
- Low noise level (<60 dB A) and Plug & Play system for easy setup
Configurations
Specifications | Values | Enquire |
---|---|---|
Model | 2-Glove Isolator (Customisable) & 4-Glove Isolator (Customisable) | |
Weight | 2-Glove Isolator: Approx. 900 kg | 4-Glove Isolator: Approx. 1100 kg | |
Power Supply | 3 x 400 V / 50 Hz / 8 kW | |
Compressed Air | H14 filtered; 6 bar [2:2:2]; 400 L/min | |
Supply/Exhaust Air | 150 – 300 m³/h; parameterisable | |
Additional Options | Integrated H₂O₂ gas generator with fully automatic material lock decontamination | Glove testing with port detection; CCTV surveillance; intercom; remote access | USB/RS232 integration; OPC UA system connection; GMP/ICH9 qualification | Sterile test pump; shelving system; fluid reservoir; RTP port solutions. |
Downloads

Pharmaceutical The Complete Solution
Download